GE OSMONICS MILLENIUMTM REVERSE OSMOSIS UNIT User Manual

Page 66

Advertising
background image

M

Millenium™ RO System by GE Osmonics

14693e - 31 Mar 03

4- 29

Routine Maintenance

4.6.2 BIOSAN™ Storage Procedure

WARNING:
It is recommended that gloves (latex/vinyl and eye protection be worn when
handling BIOSAN™.

1.

Disinfect and rinse the RO with an acceptable disinfectant.

2.

Follow the BIOSAN™ procedure in this chapter.

3.

Do not rinse the RO out after BIOSAN™ has been circulated in the RO.

4.

Disconnect the disinfect line, waste line, feed water line, power and product line.

5.

Coil the hoses and power cord, and cover the RO with a large plastic bag to
protect the RO from dust.

CAUTION:
The Millenium™ RO may be stored for up to two weeks in BIOSAN™. After two
weeks, if longer storage is needed, the Millenium™ RO may be rinsed and re-
packed with BIOSAN™.

WARNING:
After storage, the RO unit must be completely rinsed of BIOSAN™, and a bacterial
culture performed and interpreted before use. Perform AAMI analysis if it has been
one year or more since last done.

Advertising