INDICATORS


HEATING ....... This red indicator is lit when the heating

element is actually drawing current from the circuit and

heating the water. When it reaches set temperature, the

heating indicator light shuts off.


FAULT ............ If the water temperature exceeds 58

°C,

an alarm will sound and the red fault light will be

illuminated. This may indicate a unit malfunction.


ADD WATER .... This red indicator will light if there is

insufficient water in the bath. The unit will not operate.

Add water to correct this condition. This indicator will

light also when the shroud unit is lifted in up position.





7. TROUBLESHOOTING GUIDE

PROBLEM and ACTION

1) The indicator lights and displays do not come on when

the POWER key is pressed.

• Check the POWER SWITCH on the left-hand side.

• Check the outlet for 110/120 VAC 60 Hz.

• Check the fuse in the unit. Replace the fuse if necessary.

• Ensure that the POWER key is pressed for at least one

second.

• Check that there is sufficient water in the bath and that

the ADD WATER light is off.

• Check that the FAULT light is off.