2) The ADD WATER light is on and there is sufficient

water in the bath.

• With the unit unplugged and raised out of the water, check

that there is no obstruction around the float switch in the
shroud unit. The float should move freely up and down the

stem approximately half-inch.


3) The FAULT light is on.

• Check the water temperature. If the water temperature is

above 58

°C, the FAULT light will light and the unit will

not be operable.

• If this temperature is reached, it may indicate a unit

malfunction.
Contact

GenesisBPS, LLC

for further assistance.


4) Unit does not heat. HEATING light is on but actual

water temperature is not increasing.

• Remove rear cover and check the heating element for

continuity. Replace if necessary. This maintenance should
only be performed by a trained technician or

GenesisBPS,

LLC.

Contact it or your Biomedical Department.


5) Unit does not rock

• Check that the ADD WATER or FAULT lights are not

illuminated.

6) Bath moves to side when rocking or the rocking motion is

not smooth.

• Ensure that the bath is positioned on the white roller.

• The unit must be operated on a smooth and level surface.

Check the level of the work surface and make adjustments
if necessary.