GE E4H-CE User Manual

Page 11

Advertising
background image

7

Table 2.1

Alarm Messages

*

Set at factory.

ALARM

DISPLAYED

LOW INLET

PRESSURE

HI PERMEATE

PRESSURE

HI TEMP

HIGH

pH

HIGH

CONDUCTIVI-

TY

MOTOR 1

START

FAILURE

E-STOPPED

PRESSED

ACTION

Shuts down RO

machine

Shuts down RO

machine

Shuts down RO

machine

Alarm light dis-

played on pH

meter

Shuts down RO

machine

Shuts down RO

machine

Shuts down RO

machine

NOTES

Press ALARM

RESET on OIT

Press ALARM

RESET on OIT

Press ALARM

RESET on OIT

RO machine will

operate in alarm

condition

RO machine will

operate in alarm

condition

Press ALARM

RESET on OIT

and on Control

enclosure

Press ALARM

RESET on OIT on

OIT and on Control

enclosure

DISPLAY

(seconds)

12

0

0

None

None

0

None

FACTORY SET

POINTS*

Adjustable switch

[set at 0.8 bar

(12 psi)]

Adjustable switch

[set at 5.5 bar

(80 psi)]

Non-adjustable

switch [set at 38°C

(100°F)]

Adjustable switch

(set at 8.5)

Adjustable

(set at 50)

Adjustable

N/A