SECTION 4

MAINTENANCE

4.1 GENERAL

The 6300 has been designed to give optimum performance with minimum maintenance. It is only necessary
to keep the external surfaces clean and free from dust. The sample area should always be kept clean and
any accidental spillage should be wiped away immediately. To give added protection when not in use, the
unit should be disconnected from the mains supply and covered with the optional dust cover (630 028).
For longer term storage or re-shipment it is recommended that the unit be returned to the original packing
case.

NOTE:

The 6300 Monochromator is a non-serviceable unit and no attempt should be
made to repair this item. Failure to observe this recommendation will result in
the loss of any Warranty Claim on this product. In the unlikely event of the
monochromator requiring service or calibration, it is essential that the
Manufacturer or your local Distributor be contacted immediately for advice.

4.2 LIGHT SOURCE REPLACEMENT

The only routine maintenance which may be required is the replacement of the light source if this fails.
Failure should be suspected if lamp failure indicator appears on the display. This can be confirmed by
looking in the sample chamber. The Tungsten Halogen lamp is available from the Manufacturer or your
local Distributor (refer Section 5.2, Spares).

WARNING:

DISCONNECT THE UNIT FROM THE MAINS SUPPLY PRIOR TO
REPLACING THE LAMP. CARE SHOULD BE TAKEN WHEN
REMOVING THE LAMP FROM THE HOLDER. ENSURE THE LAMP IS
COOL PRIOR TO HANDLING.

1.

Access to the tungsten halogen lamp can be gained via the lamp access panel, located at the
rear of the unit (refer Section 2.4).

2.

Slacken off the thumbscrew from the lamp access panel located at the rear of the 6300.

3.

Remove the old lamp from the holder. The lamp is a plug-in fit and should be removed by
gently easing it from the holder.

IMPORTANT: WHEN FITTING THE REPLACEMENT LAMP IT IS ESSENTIAL

THAT THE GLASS ENVELOPE IS NOT TOUCHED. FINGER
MARKS WILL DAMAGE THE LAMP. SHOULD ACCIDENTAL
DAMAGE WITH FINGERMARKS OCCUR, THE SURFACE OF THE
LAMP MAY BE CLEANED USING ISO-PROPYL ALCHOHOL.

4.

Carefully remove the replacement lamp from the packaging, ensuring the glass portion of
the lamp is not touched.

5.

Insert the lamp into the holder, as illustrated, ensuring that it is fully pushed home.

10

6300/REV B/12-96