IBM Visible Range Spectrophotometer 6300 User Manual

Page 16

Advertising
background image

ASCII Fxxxx.x<CR> Sets the concentration factor to xxxx.x. For example: F1000<CR> will set the

factor to 1000.

Note <CR> is an ASCII carriage return character.
The last three commands provide an output which can readily be incorporated into most spreadsheet
software packages.

6.2 RS232 Output
The bi-directional RS232 interface is available on the rear panel 25 way D type connector.
The connections are as follows:
TXD 2

- INPUT TO 6300

RXD 3

- OUTPUT FROM 6300

RTS 4

- LINKED TO CTS

CTS 5

- LINKED TO RTS

DSR 6

- OUTPUT FROM 6300

DCD 8

- OUTPUT FRM 6300

DTR 20

- INPUT TO 6300 (must be active)

GND 7

Suggested interconnections are detailed below:
6300

IBM PC XT (25 way “D”)

TXD 2

2

TXD (From PC)

RXD 3

3

RXD (To PC)

RTS 4

4

RTS (From PC)

CTS 5

5

CTS (To PC)

DSR 6

6

DSR (To PC)

DCD 8

8

DCD (To PC)

DTR 20

20

DTR (From PC)

GND 7

7

GND

6300

IBM PC XT (9 way “D”)

TXD 2

3

TXD (From PC)

RXD 3

2

RXD (To PC)

RTS 4

7

RTS (From PC)

CTS 5

8

CTS (To PC)

DSR 6

6

DSR (To PC)

DCD 8

1

DCD (To PC)

DTR 20

4

DTR (From PC)

GND 7

5

GND

NOTE:

The Interface Cable Kit (Order Code: 542 009) can be used to implement the
above interconnections.

Recorder Output
This is available via the 4mm rear panel sockets. The level is proportional to the displayed reading, de-
pending on the measurement mode:

Transmission

1mV per 0.1%T

Absorbance

1mV per 0.001ABS

Concentration

1mV per concentration unit

13