MULTI FUNCTION CHARGER PORT

ƽ WARNING

Set-up/programming of the Electronic

Control Unit is to be performed only by

a qualified technician. The final adjust-

ments of the controller may affect other

activities of the wheelchair. Damage to

the equipment could occur under these

circumstances. If unqualified individuals

perform any work on these units, the

warranty is void.

FIGURE 5.2 - MULTI FUNCTION

CHARGER PORT

The multi function charger port is located at the front of the joystick housing. This
provides easy access for charging the wheelchair batteries. This port also serves as the
Remote Programmer Communication connection.

RESETTING THE CIRCUIT BREAKER

ƽ WARNING

NEVER defeat or bypass the circuit breaker. ONLY replace with a circuit breaker of

the same rating.

NOTE: For this procedure, refer to FIGURE 5.3.

1. Remove the battery pack from the wheelchair base. Refer to Removing/Installing the

Battery Pack on page 35.

2. Open the battery pack by performing the following -

A. Grasp the side flaps and pull up.

B. Grasp the front flap and pull up.

3. To reset - press the circuit breaker reset

button located inside the protector ring
on top of circuit breaker.

4. Close the battery pack by performing

the following -

A. Fold down the front flap and press

to engage the hook and loop straps.

B. Fold down the side flaps and press

to engage the hook and loop straps.

NOTE: Ensure that the hook and loop straps
are securely fastened before lifting the battery
pack.

FIGURE 5.3 - RESETTING THE

CIRCUIT BREAKER

Multi-Function

Charger Port

FRONT VIEW OF JOYSTICK

Circuit Breaker

Reset Button

Side Flap

Side Flap

Front Flap

Protector Ring