8.2.3.1 Removing the batteries

Requirements:

•

Wrench 11 mm

•

Remove the seat.

•

Remove the battery and motor compartment cover.

•

Disconnect the main cable harness.

•

Open the belts that hold the batteries in place (1).

•

Loosen battery clamp of the blue cable at the negative pole
of the battery with the wrench and remove cable.

•

Loosen the battery clamp of the red cable at the positive
pole of the battery with the wrench and remove cable.

•

Repeat procedure on the other side.

NOTE
Fitting of new batteries is done in reverse order.