Procedures for Charging.
24V 10A charger

The 24V 10A charger has one

bi-coloured orange/green indicator

to show the charger operation

(Fig. 29).
1 Plug into the three pin charging

point under the fornt of the

control box (Fig. 28).

2 Connect the charger to the

mains supply by means of the

mains plug and switch on.

3 At the start of charging the

charger will undergo a self test

procedure, this is indicated by the

indicator flashing, alternating

orange/green (Fig. 29).

4 When the test is complete and

there have been no faults found,

charging will commence. This is

indicated by a flashing orange

only light.

5 On return of 90% or the bulk of

the charge a steady orange light

will be visible.

6 When a steady green light is

visible the battery is fully charged.

7 An automatic switch to a float

voltage will occur after a further

period of time, and the charger

should be left connected to the

wheelchair until it is required for

use.

8 No light indicator will show

unless the charging plugs are

connected.

9 Always switch off at the mains

before disconnecting the
batteries.