Appendix D: System Electromagnetic Emissions and Immunity Declarations

Table 204 – Guidance and manufacturer’s declaration – electromagnetic immunity – for
EQUIPMENT and SYSTEMS that are not LIFE-SUPPORTING (see 6.8.3.201 b))

Guidance and manufacturer’s declaration – electromagnetic immunity

The MC

2

Series is intended for use in the electromagnetic environment specified below. The

customer or the user of the MC

2

Series should assure that it is used in such an environment.

Immunity test

IEC 60601 test
level

Compliance
level

Electromagnetic immunity - guidance

Conducted
RF IEC
61000-4-6

Radiated RF
IEC 61000-4-
3

3 Vrms

150 kHz to 80
MHz

3 V/m

80 MHz to 2.5
GHz

Portable and mobile RF communications
equipment should be used no closer to any
part of the MC

2

Series, including cables, than

the recommended separation distance
calculated from the equation applicable to
the frequency of the transmitter.

Recommended separation distance

d = [3.5/V

1

]e

P

d = [3.5/E

1

]e

P 80 MHz to 800 MHz

3 V

d = [7/E

1

]e

P 800 MHz to 2.5 GHz

where P is the maximum output power rating
of the transmitter in watts (W) according to
the transmitter manufacturer and

d is the

recommended separation distance in meters
(m).

Field strengths from fixed RF transmitters, as
determined by an electromagnetic site
survey,

a

should be less than the compliance

level in each frequency range.

b

3 V/m

Interference may occur in the vicinity of
equipment marked with the following symbol:

NOTE 1 At 80 MHz and 800 MHz, the higher frequency range applies

NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is
affected by absorption and reflection from structures, objects and people.

a

Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless)

telephones and land mobile radios, amateur radio, AM and FM radio broadcast and TV broadcast
cannot be predicted theoretically with accuracy. To assess the electromagnetic environment due
to fixed RF transmitters, an electromagnetic site survey should be considered. If the measured
field strength in the location in which the MC

2

Series is used exceeds the applicable RF

compliance level above, the MC

2

Series should be observed to verify normal operation. If

abnormal performance is observed, additional measures may be necessary, such as reorienting
or relocating the MC