Stinger MC POWER SYSTEM MC2 User Manual

Page 61

Advertising
background image

MC

2

Power Systems User's Guide

Appendix D: System Electromagnetic Emissions and Immunity Declarations

1110227 Revision 2

November 2006

49

Table 206 – Recommended separation distances between portable and mobile RF
communications equipment and the EQUIPMENT or SYSTEM – for EQUIPMENT and
SYSTEMS that are not LIFE-SUPPORTING (see 6.8.3.201 b)

Recommended separation distances between

portable and mobile RF communications equipment and the MC

2

Series

The MC

2

Series is intended for use in an electromagnetic environment in which RF disturbances

are controlled. The customer or the user of the MC

2

Series can help prevent electromagnetic

interference by maintaining a minimum distance between portable and mobile RF
communications equipment (transmitters) and the MC

2

Series as recommended below, according

to the maximum output power of the communications equipment.

Separation distance according to frequency of transmitter

m

Rated maximum output
power of transmitter

W

150 kHz to 80
MHz

d = [3.5/V

1

]eP

80 MHz to 800 MHz

d = [3.5/E

1

]eP

800 MHz to 2.5 GHz

d = [7/E

1

]eP

0.01 0.12

0.12

0.23

0.1 0.38

0.38

0.73

1 1.2

1.2

2.3

10 3.8

3.8

7.3

100 12

12

23

For transmitters rated at a maximum output power not listed above, the recommended separation
distance d in meters (m) can be estimated using the equation applicable to the frequency of the
transmitter, where P is the maximum output power rating of the transmitter in watts (W) according
to the transmitter manufacturer.

NOTE 1 At 80 MHz and 800 MHz, the separation distance for the higher frequency range
applies.

NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is
affected by absorption and reflection from structures, objects and people.

Advertising