Lakeshore Learning Materials 642 User Manual

Page 104

Advertising
background image

Lake Shore Model 642 Electromagnet Power Supply User’s Manual

7-12

Service

7.10.3 Auxiliary

Connector

The Auxiliary Connector provides connections for three functions:
1. Emergency Stop – This normally-closed circuit turns off power to the Model 642 just as if the OFF (O) button was

pressed on the front panel when opened. Normal operation requires a closed connection between the pins.

2. Fault Relay – The fault relay can be used to communicate the presence of a Model 642 fault to external equipment.

The relay follows the operation of the Fault light on the front panel. Both normally-open and normally-closed
configurations are provided. The contacts are electrically isolated from the Model 642 chassis.

3. Remote Enable – These contacts are similar in function to the flow switch inputs. Normal (enabled) operation

requires a closed connection between the pins.

1

8

Pin Name

1 Emergency

Stop

A

2

Emergency Stop B

3 Chassis-common

4 Fault-NO
5 Fault-Com
6 Fault-NC
7 Chassis-common
8 Remote

Enable

Figure 7-9. Auxiliary Connector Details