Digital i/o – Philips Inca 311 User Manual

Page 9

Advertising
background image

2.3.3 Digital

I/O

The Digital Input and Output connectors give the user
the possibility to connect and control a number of devices.
For that purpose 6 output and 6 input lines are available.
These output and input lines are optically isolated from the
Inca (see explanation: 2.3.4 How to use the isolated
input/output).
Input 6 differs from the other inputs in a way that it has a
special purpose. Input 6 can be used as an interrupt input
either level or edge sensitive.

All inputs and outputs are TTL level compatible.
• input current 6.3 mA < Ion < 10 mA

• Pmax 20 mW

• output current 0.9 mA < Ion < 18 mA, Vce max 40 volt








2005-12-12

Page 7

Figure 2-3: Digital input and output connections

Table 2-1: LED control by Rhapsody software
LED 1

green or yellow

LED 2

green or yellow

LED 3

green or yellow

LED 4

green (only!)


DIGITAL I/O

PIN

INPUT

Input 4(p)

Input 1(p)

Input 2(p)

Input 6(p)

Input 3(p)

Return input 6(n)

Input 5(p)

8

7

6

5

4

3

2

1

Return input 1..5(n)

PIN

OUTPUT

Output 4(p)

Return output 4..6(n)

Output 6(p)

Output 2(p)

Output 5(p)

Output 1(p)

Return output 1..3(n)

8

7

6

5

4

3

2

1

Output 3(p)

(p/n see: 2.3.4 How to use the isolated input/output)


Note: The common grounds of the outputs are clustered in
two groups of three outputs. Five input common grounds
are clustered.

LED 1

LED 2

1

Input

Output

1

LED 3

LED 4

Advertising