Defining a calibration template, Figure 6-1. defining a calibration grid, Defining a calibration template -2 – National Instruments IMAQTM User Manual

Page 87

Advertising
background image

Chapter 6

Calibrating Images

IMAQ Vision for Visual Basic User Manual

6-2

ni.com

Refer to Chapter 5,

Performing Machine Vision Tasks

, for more

information about applying calibration information before making
measurements.

Defining a Calibration Template

You can define a calibration template by supplying an image of a grid or
providing a list of pixel coordinates and their corresponding real-world
coordinates. This section discusses the grid method in detail.

A calibration template is a user-defined grid of circular dots. As shown in
Figure 6-1, the grid has constant spacings in the x and y directions. You can
use any grid, but follow these guidelines for the best results:

The displacement in the x and y directions must equal (dx = dy).

The dots must cover the appropriate portion of the working area.

The radius of the dots must be 6–10 pixels.

The center-to-center distance between dots must range from
18 to 32 pixels, as shown in Figure 6-1.

The minimum distance between the edges of the dots must be 6 pixels,
as shown in Figure 6-1.

Figure 6-1. Defining a Calibration Grid

Note

You can use the calibration grid installed with IMAQ Vision at Start»Programs»

National Instruments»Vision»Documentation»Calibration Grid. The dots have radii
of 2 mm and center-to-center distances of 1 cm. Depending on the printer, these
measurements may change by a fraction of a millimeter. You can purchase highly
accurate calibration grids from optics suppliers, such as Edmund Industrial Optics.

1

Center-to-Center Distance

2

Center of Grid Dots

3

Distance Between Dot Edges

dy

dx

1

3

2

Advertising