Replacing Fuses

To replace a blown fuse on the accessory, complete the following steps.

1.

Unplug the accessory power.

2.

Remove all signal wires and cables from the accessory.

3.

Unscrew the two screws from a side panel with a Phillips head screwdriver and remove the side
panel as shown in Figure 4.

Figure 4.

Removing the Circuit Board

4.

Slide the circuit board out.

5.

Replace the blown fuse while referring to Figure 1 for the fuse location.

Spare fuses can be found next to each fuse location as shown in Figure 1. You also can order
additional fuses from Littelfuse at

www.littelfuse.com

or from electronics distributors.

Fuse

Littelfuse Part Number

Description

Power

0448002

2A, 125V lead-free very fast-acting NANO

2

subminiature

ceramic surface mount fuse, 6.10

× 2.69 mm

Iso Out 0,
Iso Out 1,
RS-232

0448.200

200mA, 125V lead-free very fast-acting NANO

2

subminiature ceramic surface mount fuse, 6.10

× 2.69 mm

NA