Signal Descriptions

Channel Input Signal Connections

The signal terminals for the positive input channels are located in
column C of the connector. The signal terminal for each corresponding
negative input channel is located in column B of the connector. Each
input goes to a separate filter and amplifier that is multiplexed to the
module output buffer. The terminal block temperature sensor output—
connected to pins A3 and A4 (CJSENSOR)—is also filtered and
multiplexed to the module output buffer.

The differential input signal range of a VXI-SC-1102/B/C input channel
is

±

10 V/G, where G is the gain selected on the VXI-SC-1102/B/C input

channel. This differential input range is the maximum measurable
voltage difference between the positive and negative channel inputs.
The common-mode input signal range of a VXI-SC-1102/B/C input

Table 3-1. Front Connector Signal Descriptions

Pin

Signal Name

Description

A1

+5 V

+5 VDC Source—Powers the
temperature sensor on the terminal
block. 0.2 mA of source not
protected.

A2, A5, A16,
A24, A32

CGND

Chassis Ground—Tied to the
VXIbus chassis.

A3, A4

CJSENSOR

Cold-Junction Temperature
Sensor Input—Connects to the
temperature sensor of the terminal
block.

B1–B32

CH31-
through
CH0-

Negative Input Channels 31–0
—Negative side of differential
input channels.

C1–C32

CH31+
through
CH0+

Positive Input Channels 31–0
—Positive side of differential
input channels.

All other pins are not connected.