Cold-junction sensor connection, Figure 3-3. floating signal connection referenced, Cold-junction sensor connection -6 – National Instruments VXI-SC-1102 User Manual

Page 27: Figure 3-3

Advertising
background image

Chapter 3

Signal Connections

VXI-SC-1102/B/C User Manual

3-6

©

National Instruments Corporation

Figure 3-3. Floating Signal Connection Referenced to Chassis Ground

Cold-Junction Sensor Connection

Pins A3 and A4 (CJSENSOR) connect the temperature sensor located
on the National Instruments VXI-TB Series terminal block to the
VXI-SC-1102/B/C. The CJSENSOR signal is measured relative to
CGND. Pins A3 and A4 are connected together within the
VXI-SC-1102/B/C so the position of the MTEMP/DTEMP switch on
the VXI-TB Series terminal block does not matter. The input is
overvoltage-protected to 15 VDC with the power on and off.

Caution:

Exceeding the overvoltage protection on the CJSENSOR input can damage
the VXI-SC-1102/B/C submodule, the VXIbus, and the VXI-MIO board.
National Instruments is

NOT

liable for any damages or injuries resulting

from such signal connections.

V

s

-

+

CH0 +

Front Signal

Connector

+

-

CH0 -

+

-

CH1 +

CH1 -

VXI-SC-1102/B/C

CGND

Reference to

Chassis Ground

Floating

Signal

Make this connection to

ground reference the signal.

!