Option 2: thermal disinfection – ResMed Non-Vented Full Face Mask Mirage Series 2 User Manual

Page 4

Advertising
background image

Clinical Use Only

3

En

gli

sh

Option 2: Thermal Disinfection

Pre-disinfection

1 Disassemble the mask.
2 Clean the individual parts of the mask with a soft bristle brush for one

minute while soaking them in detergent (eg, Neodisher

Medizyme or

Alconox). Pay particular attention to all crevices and cavities.

3 Rinse the components twice by agitating them vigorously in drinking quality

water and allow them to air dry out of direct sunlight.

ResMed has tested the detergents Neodisher Medizyme (concentration at
2% for 30 minutes) and Alconox (diluted at 1%).

Disinfection

Using a certified hot water disinfection system, soak the disinfectable mask
components using

one of the temperature-time combinations listed in the

following sections.

European Standards

Based on prEN ISO 15883-1, the mask has been tested at the following
temperature-time combinations:

70°C for 100 minutes

75°C for 30 minutes

80°C for 10 minutes

90°C for 1 minute.

Regional Requirements

Due to specific regional requirements, the mask has also been tested for
disinfection at the following temperature-time combination:

93°C for 10 minutes.

Testing by ResMed has shown that the mask components can withstand
15 cycles of thermal disinfection using the above temperature-time
combinations. The device used for testing was Branson

LTH 1828-24 hot

water bath.

Advertising
This manual is related to the following products: