Omega Speaker Systems OS530LE User Manual

Page 22

Advertising
background image

2-7

Using the Handheld Infrared Thermometer

2

• Static Surface Scan

– Measures the temperature across a static

surface:

1.

Aim the thermometer at a starting point and pull the trigger.
Press the

key to lock the trigger.

2.

If necessary, adjust the emissivity.

3.

Slowly move the thermometer so that the line of sight sweeps

across the surface. The thermometer measures the temperature

at each point on the surface.

4.

To record the temperature profile across the surface, connect

the IR thermometer to a strip chart recorder. Refer to Figure

2-11 for details. The IR thermometer provides an analog output

of 1mV/degree. (0.5 mV/Deg on OS524E)

5.

After all the data has been taken, press the

key to unlock

the trigger.

Figure 2-11 Recorder Hookup

• Moving Surface Scan

- Measures the Temperature of Points on a

Moving Surface:

1.

Mount the thermometer on a camera tripod and aim at a fixed
point on the moving surface.

2.

Pull the trigger and press the

key to lock the trigger.

3.

If necessary, adjust the emissivity. The thermometer is now set
up for measuring the temperature of a moving surface.

4.

To record the temperature profile of the moving surface,
connect the IR thermometer to a strip chart recorder. Refer to
Figure 2-11 for details.

5.

After all data is taken, press the

key to unlock the trigger.

• Fixed Point Monitoring Over Time

- Monitors the temperature at

a fixed point over time:

It is recommended that you use the ac adapter for long
term measurement of temperature.

LOCK

LOCK

LOCK

LOCK

NOTE

To Strip Chart

Recorder

Analog

Cable

NOTE

Center hole is the