Omega Speaker Systems OS530LE User Manual

Page 34

Advertising
background image

2-19

Using the Handheld Infrared Thermometer

2

2.3.11 Using Ambient Target Temperature

Compensation

(OS533E, OS534E, OS523E, OS524E)

Use the Ambient Target Temperature Compensation

(AMB)

Display Mode when high accuracy readings under both

of these conditions are required:

• The target has a low emissivity.
• The ambient temperature around the target is much

higher than the ambient temperature around the
infrared thermometer.

To set and activate the Ambient Target Temperature
Compensation Mode:

1.

Pull the trigger and press the

key to lock the

trigger. Set the emissivity to 1.0 (refer to Section 2.3.1).

2.

Press and hold the

key until the Average Display

Mode (AVG) appears.

3.

Slowly move the thermometer so that the line of sight
sweeps across the area surrounding the target. The
thermometer measures the temperature at each point on
the surrounding area.

4.

Read the average temperature value from the upper
display and record it here __________.

5.

Press and hold the

key until the Ambient

Temperature Display Mode (AMB) appears.

6.

Set the AMB temperature found in Step 4 by pressing the

key or the

key.

7.

Press the

key to enable the ambient target

temperature compensation. The

icon appears

on the display.

To disable this mode, press the

key again.

The icon

disappears.

F

C

LOCK

LOCK

FUNC

LOCK

FUNC

NOTE

°F

ATC