Omega Vehicle Security 450 User Manual

Page 19

Advertising
background image

16

NOTE: If you want to cancel the calibration
mode press OFF button in front of the unit.

For T/C type

K set simulator @ 1320

°

C

J set simulator @ 960

°

C

T set simulator @ 350

°

C

E set simulator @ 950

°

C

Press the switch one more time and wait for a few
seconds until display switches to 0.0 set simulator
to 0

°

C. Press switch one more time. Unit should

shut-off.

Step 4. Now, turn the unit on from front panel.
Make sure the unit in degree C mode. Verify that
the unit is within specification throughout its range.

Step 5. If unit does not meet specification as 0

°

C

then repeat calibration procedure,
Instead of setting simulator as 0

°

C, set it at the

temperature indicated by the unit in measurement
mode as 0

°

C. For example in measure mode, unit

may display .2

°

C with simulator as 0

°

C. During