polrity).

j. The voltage measured at the “3.7V” test-

point should be 3.007 times larger than
the voltage measured at the “-1.23V” test-
point. Adjust Rc (which varies the +3.7V
line to attain the proper ratio.

k. Switch the indicator to the 1

°

range.

l. Adjust the voltage calibrator for an output of

+53.782mV for type K; +57.942mV for type J;
+75.983mV for type E; +18.824mV for type T;

m. Adjust Ra for an indicator display of +995

°

C

for type E, or 1000

°

C for type J, or +1340

°

C

for type K, or +367

°

C for type T.

n. Reassemble the indicator.

o. Perform the zero calibration.