Toshiba LF610 User Manual

Setting parameters flow  setting menu

Advertising
background image

Electromagnetic Flowmeter LF610 Series
Operation Guide

Setting Parameters Flow

Setting Menu

Items

Screen Display

Setting or selecting value

A

A

DETECTOR

1 Exicitation Current

EXC CUR

0.0000A

0.2500A (0.0001A each)

*1

2 Meter size

SIZE

0.1

36 inch , 2.5mm 900mm

*1

3 Exicitation Frequency

EXC FREQ

6, 12, 24 Hz

4 Flow Direction

FLOW DIR

NORMAL , SWITCH

5 Measurement Mode

EXIT

B

B

DISPLAY

1 Display 1

DSPL1

First Unit : Volume, Others

%, m

3

, L, mL, bbl, gal, m/s, ft/s, COUNT, RANGE, pt, qt,

CUSTOM, GRAPH (*3)

Second Unit : Time unit

/s, /min, /h, /d

Third unit : Flow direction code

B (Automatic selection bi-directionflow)

F (Fixed forward flow), R (Fixed reverse flow)

D (Difference)

*2

*3

2 Display 2

DSPL2

3 Custom Coefficient

CS VAL

0

99999999

4 Custom Unit

CS UNIT

7 Characters

5 Measurement Mode

EXIT

C

C

RANGE

1 Range Type

R TYPE

SINGLE, 4F-0R, 2F-2R, EXT 2F-0R, EXT 2F-2R

*4

2 Range 1

R1

Volume unit

m

3

, L, mL, bbl, gal, pt, qt

Time unit

/s, /min, /h, /d

Velocity unit

m/s, ft/s

3 Range 2

R2

4 Range 3

R3

5 Range 4

R4

6 Range Hysteresis

R HYS

0.0

25.0 % (0.1% each)

7 Measurement Mode

EXIT

D

D

FILTER

1 Damping Constant

DAMPING

00.0, 00.5, 01.0

60.0 s

2 Rate-of-change Limit

LIM RATE

0.0

30.0 % (0.1% each)

3 Control Limit Time

LIM TIME

00

20 s (1s each)

4 Measurement Mode

EXIT

E

E

LOW CUT

1 Low Cutoff

CUT VAL

0.0

10.0 % (0.1% each)

2 Display Low Cutoff

DSPL SET

LINEAR, LOW CUT

3 Measurement Mode

EXIT

F

F

ZERO

1 Zero Adjustment

ZERO ADJ

Hold down the

key

2 Manual Zero

MANUAL

±10% of 10m/s-maximum range (0.1% each)

*5

3 Measurement Mode

EXIT

G

G

4-20mA

1 4-20mA Alam Output

ALM 4-20

UNDER 3mA, 4mA, HOLD , OVER 24mA

2 Output Low Limit

LOW LIM

4.0mA, 3.2mA, 2.4mA

*6

3 Measurement Mode

EXIT

H

H

DO

1 Digital Output 1

DO1 FUNC

NO USE, HIGH ALM, HH ALM, LOW ALM, LL ALM

RNG SIG1, RNG SIG2, PRESET C, CONV ALM

EMPTY ALM, PLS OUT, PLS FRD, PLS REV

*7

2 Digital Output 2 (*13)

DO2 FUNC

3 Digital Output 1 Active Status

DO1 STAT

NormOPEN, NormCLOSE

*8

4 Digital Output 2 Active Status (*13)

DO2 STAT

5 Measurement Mode

EXIT

I

I

DI

1 Digital Input (*13)

DI FUNC

NO USE, C ST/SP, C RS/ST

RNG SW, ZERO ADJ, FIX OUT

2 Digital Input Detective Level (*13)

DET LVL

H LEVEL, L LEVEL

*9

3 Measurement Mode

EXIT

J

J

CNT/PLS

1 Counting Rate

CNT RATE

Units : m

3

, L, mL, bbl, gal, pt, qt

Value : within the range 3.6

36000000 pulse/h

(0.001

10,000 pps)

2 Plulse Mode

PLS MODE

AUTO, MANUAL

3 Pulse Width

PLS WID

0.3

500.0ms (0.1ms each)

or Less than 40% of the pulse rate for 100% flow rate output

4 Measurement Mode

EXIT

K

K

PRESET C

1 Preset Count

PRST VAL

0

99999999 count (1 count each)

2 Preset Function

OUT MODE

HOLD, 50ms PLS, 500ms PLS

*10

3 Measurement Mode

EXIT

L

L

H/L ALM1

1 High Alarm Set

H SET

ON, OFF

2 High Alarm Value

H VAL

-10

110 % (0.1% each)

3 Low Alarm Set

L SET

ON, OFF

4 Low Alarm Value

L VAL

-10

110 % (0.1% each)

5 Measurement Mode

EXIT

M

M

H/L ALM2

1 High High Alarm Set

HH SET

ON, OFF

2 High High Alarm Value

HH VAL

-10

110 % (0.1% each)

3 Low Low Alarm Set

LL SET

ON, OFF

4 Low Low Alarm Value

LL VAL

-10

110 % (0.1% each)

5 Measurement Mode

EXIT

N

N

SELF CHK

1 Empty Pipe Alarm

EMPTY

OFF, NORMAL, SENS, SENS-H

*11

2 Self-diagnosis function

SELF CHK

ON, OFF

3 Alarm Output Preset

CONV ALM

CONV ONLY, WITH EMP

4 Measurement Mode

EXIT

O

O

FIX OUT

1 Fixed-value Output

FIX SET

ON, OFF

*12

2 Fixed-current Output

CUR VAL

2.4

24.0 mA (0.1mA each)

3 Fixed-pulse Output

PLS VAL

0

10000 pps (1pps each)

4 Measurement Mode

EXIT

P

P

OTHERS

1 Password

PASSWORD

000

999 (1 each)

2 LCD Density Adjustment

LCD ADJ

1(Light)

5 (Dark)

3 Swich Position

SW POSN

TOP, BOTTOM, LEFT, RIGHT

4 Measurement Mode

EXIT

Q

Q

COMM

1 PROFIBUS Communication

PROFIBUS

*13

2 MODBUS Communication

MODBUS

*13

3 Measurement Mode

EXIT

R

R

CAL

1 0% Flow Rate Calibration

CAL 0%

2 50% Flow Rate Calibration

CAL 50%

3 100% Flow Rate Calibration

CAL 100%

4 Checking the Excitation Current Value EXC DSPL

5 Measurement Mode

EXIT

*1 This value is factory adjusted when shipped.

*2

In case of choosing COUNT or RANGE

COUNT : displays totalized flow counts (8 digits) without a unit.

RANGE : displays the range number (1 to 4)

*3 Only display2

*4

Range type

Description

1 : SINGLE

Single range

2 : 4F-0R

Unidirectional flw , automatic selection of multiple ranges

3 : 2F-2R

Bidirectional flows , automatic selestion of multiple ranges

4 : EXT 2F-0R

Unidirectional flows , multiple ranges selected by exterminal single

5 : EXT 2F-2R

Bidirectional flows , multiple ranges selected by exterminal single

*5

(1) In case of the zero adjustment (Menu No.F1),Zero offset is automatically cleared to Zero.

(2) Calculate the zero offset value with the follwing equation:

Zero offset value (%) = {(Actual flow rate) - (LF610 measured value)}

(Sample)

Measured condition

Flow rate

% in measured span

Actual flow rate obtained

from other instrument

10.0 m

3

/min

50%

LF610 measured value

10.5 m

3

/min

52.5%

Zero offset

-

-2.5%

If zero offset is set to -2.5% , LF610 will output 50.0% flow rate instead of 52.5%

*6

4-20mA Alarm Output

Output Status

0 : UNDER 3mA

Under 3mA output

1 : 4mA

4mA output

2 : HOLD

Measured data hold

3 : OVER 24mA

Over 24mA output

*7

DO1,DO2 items

Digital output function

0 : NO USE

Not used

1 : HIGH ALM

High limit alarm output

2 : HH ALM

High high limit alarm output

3 : LOW ALM

Low limit alarm output

4 : LL ALM

Low low limit alarm output

5 : RNG SIG1

Multi - range output No.1

6 : RNG SIG2

Multi - range output No.2

7 : PRESET C

Preset point output

8 : CONV ALM

Converter failure alarm output

9 : EMPTY ALM

Empty pipe alarm output

10 : PLS OUT

Pulse output (bidrectional flow)

11 : PLS FRD

Pulse output (fixed forward flow)

12 : PLS REV

Pulse output (fixed reverse flow)

*8

DO1/DO2 Active Status

Alarm - output Status

0 : NormCLOSE

Normal close

1 : NormOPEN

Normal open

*9

DI Detective Level

Detective Level

0 : L LEVEL

L level

1 : H LEVEL

H level

*10

Preset Function

Output Status

0 : HOLD

Output Status Level Hold

1 : 50ms PLS

Pulse out (pulse width 50ms)

2 : 500ms PLS

Pulse out (pulse width 500ms)

*11

Empty Pipe Alarm

Digital output function

0 : OFF

Not used

1 : NORMAL

Used and Low-sensitive

2 : SENS

Used and Middle-sensitive

2 : SENS-H

Used and High-sensitive

*12

When this function is set to ON , the conditions are follwing.

Items

Conditions

Current output

User - set current output

Pulse output

Pulse output with a user - set counting rate

Digital outputs

Previous status is retained (excluding pulse output).

Data indicating

Instaneous flow rates and flow velocity (no totalization).

*13 These functions are option.