0

The time specified in this command must be between 0 and 6.5535 seconds.

ABORt, ROUTe:SCAN
TRIGger[:SEQuence]:SOURce
TRIGger[:SEQuence][:IMMediate]
TRIGger[:SEQuence]:DELay

This command specifies the time to wait after a trigger event is detectedbefore
opening the current relay in a scan list. A trigger event is enabled with teh
TRIGger:[SEQuence]:SOURce command.

Command

Response

route:scan

(@m1(1,2),m2(1:64))

Define a scan sequence consisting of relays 1 and 2 on relay
module m1 and relays 1 through 64 on module m2.

trigger:sequence:count

2

Define the number of times to sequence through the scan list
before setting the state of the trigger subsystem back to the
idle state.

trigger:sequence:delay

0.5

Wait 0.5 seconds after a trigger event is detected before open-
ing the current relay in a scan list and closing the next relay in
the list.

trigger:sequence:

source ttltrg2

Sequence to the next relay in a scan list when VXI TTL trigger
2 is asserted.

initiate:immediate

Initiate the scan sequence. Sequence through the entire scan
list two times. Wait 0.5 seconds after VXI TTL trigger 2 is as-
serted before sequencing to the next relay in the list.