Tektronix 070-9180-01 User Manual

Page 87

Advertising
background image

IEEE-488.2 Common Commands

Option 01 VXI Interface Module User Manual

3–63

exception of bit 6) are also returned in response to a VXI Read STB command.
Bit 6 is read destructively in response to a VXI Read STB command. Bit 6 is not
read destructively is response to a *STB? query. The Status Byte register is
encoded as follows:

Bit No.

Meaning

bit 0 (lsb)

Not used

bit 1

Not used

bit 2

Set when the Error/Event queue contains one or more error messages.

bit 3

Not used

bit 4

Set when the Output buffer contains one or more query responses.

bit 5

Set when the contents of the Standard Event Status register ANDed
with the contents of the Standard Event Status Enable register
produces a non-zero result.

bit 6

Set when the contents of bits 0 through 5 and bit 7 of the Status Byte
register ANDed with the contents of the Service Request Enable
register produce a non-zero result.

If bit 6 becomes set, the Option 01 will generate a VXI Request True
interrupt on the VME interrupt level selected by its interrupt level select
switch. If bit 6 becomes set and the Status Byte register is read using a
VXI Read STB command, bit 6 will be reported as being set. In
response to subsequent VXI Read STB commands, bit 6 will be
reported as being cleared.

If bit 6 becomes set and the Status Byte register is read using a *STB?
query, bit 6 will be reported as being set. Bit 6 will continue to be
reported as set in reponse to *STB queries until the contents of bits 0
through 5 and bit 7 of the Status Byte register ANDed with the contents
of the Service Request Enable register is equal to 0.

bit 7

Not used.

Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com

Advertising