Quickstart Guide

Teledyne Analytical Instruments

15

Auto Mode Zero Calibration

•

Press Zero to enter the zero function mode.

•

Use the ▲▼ arrow keys to toggle between

AUTO

and

MAN

zero settling. Stop when

AUTO

appears blinking.

Zero: Settling: AUTO

<ENT> To Begin

•

Press Enter to begin zeroing.

#### PPM Zero

Slope=#### ppm/s

.
.
.

#### PPM Zero

4 Left=### ppm/s

The zeroing process will automatically conclude when the output is within the acceptable

range for a good zero. Then the analyzer automatically returns to the Analyze mode.

Manual Mode Zeroing

•

Press Zero to enter the Zero function.

•

Use the ▲▼ keys to toggle between

AUTO

and

MAN

zero settling. Stop when

MAN

appears, blinking, on the display.

Zero: Settling: Man

<ENT> To Begin

•

Press Enter to begin the zero calibration.

#### ppm Zero

Slope=#### ppm/s

Once zero settling is completed, the instrument automatically returns to the Analyze mode.