Teledyne Trace Oxygen Analyzer 3000TA User Manual

Page 16

Advertising
background image

Quickstart Guide

Trace Oxygen Analyzer


Teledyne Analytical Instruments

16

Span Calibration

Make sure an appropriate span gas is connected to the sample port.


Auto Mode Spanning:

Press Span to enter the span function.

Use the arrow keys to toggle between

AUTO

and

MAN

span settling. Stop when

AUTO

appears, blinking on the display.

Span: Settling: AUTO

<ENT> For Next

Press Enter.

Calib. Holding time

Cal hold: 5 min

Press Enter to move to the next screen.

Span Val: 000008.00

<ENT>Span <UP>Mod #

Use the ▲▼ arrow keys to enter the oxygen-concentration mode. The ◄► arrows
choose the digit, and the ▲▼ arrows choose the value of the digit. When you have
finished typing in the concentration of the span gas you are using, press Enter to
begin the span calibration.

#### ppm Span

Slope=#### ppm/s

The instrument automatically returns to the Analyze mode at the conclusion of the auto span

cycle..

Manual Mode Spanning:

Press Span to start the Span function.

Span: Settling:MAN

<ENT> For Next

Use the ▲▼ keys to toggle between

AUTO

and

MAN

span settling. Stop when

MAN

appears, blinking, on the display. Press Enter to move to the next screen.