Preparations before Starting Observation

12

Page

2-3-2 VS120-L100

The following additional precautions must be observed to prevent

automatic transfer errors from occurring:

Keep slide glass specimens free of moisture or water.

Clean the reverse face of a slide glass specimen to remove dust or

stains.

Do not deliver the cassette being tilted. The slide glass specimen in

the cassette may drop out.

1. Hold and lift the cassette cover open/close handle (1) of the slide loader to open

the cassette cover.

2. Take two cassettes out by holding the knob (2) at the top of the cassette.

3. Set the cassette on a level surface with the cassette lid (3) facing up.

4. Orient a slide glass specimen (4) such that the label (5) faces the lid and the cover

glass faces the knob. Insert the slide glass specimen (4) oriented this way into the

cassette.

(6) Be careful not to insert one slide glass specimen diagonally into

two different slots.

(7) Be careful not to insert more than one slide glass specimen into a

single slot.

Slide glass specimens are transferred into the microscope system in the

order indicated by the arrow.

5. Gently set the cassette loaded with slide glass specimens upright. Hold the knob

while taking care not to tilt the cassette, hold it so the lid faces left, and then put

the cassette back into place inside the slide loader.

6. Check that the CASSETTE1 (or CASSETTE 2) status indicating LED on the front

face of the slide loader lights up green.

7. Close the cassette cover on the slide loader.

TIP

NOTE

NOTE

NOTE

(1)

(4)

(5)

(3)

(2)

(6)

(7)