Wordclock Connections & Termination

For correct operation it is essential that the wordclock cabling be terminated correctly. The
03D has a wordclock termination ON/OFF (75

Ω

) switch on the rear panel. Wordclock is a

TTL signal, and IN and OUT connections use BNC connectors. Three wordclock
distribution examples are shown below. Note the 75

Ω

wordclock terminator switch

settings.

1. Parallel Distribution Box

In this example, a parallel distribution box is used to distribute the wordclock signal among
devices. All wordclock terminator switches are set to ON.

2. Using BNC T-bar Connectors

This example is similar to the above except that T-bar connectors are used. In this system,
only the last device’s wordclock terminator switch is set to ON.

3. Daisy Chain Distribution

In this example, the wordclock master is a digital multitrack recorder. Both 03D wordclock
terminator switches are set to ON. This method of wordclock distribution is not
recommended for large systems.

03D-A

75

Ω

Switch=ON

03D-B

75

Ω

Switch=ON

Other

Device-A

75

Ω

Switch=ON

Other

Device-B

75

Ω

Switch=ON

WORDCLOCK
OUT

WORDCLOCK IN

WORDCLOCK IN

WORDCLOCK IN

WORDCLOCK IN

Distribution box

03D-A

75

Ω

Switch=OFF

03D-B

75

Ω

Switch=OFF

Other

Device-A

75

Ω

Switch=OFF

Other

Device-B

75

Ω

Switch=ON

WORDCLOCK
OUT

WORDCLOCK IN

WORDCLOCK IN

WORDCLOCK IN

WORDCLOCK IN

03D-A

75

Ω

Switch=ON

IN

OUT

03D-B

75

Ω

Switch=ON

WORDCLOCK
OUT

WORDCLOCK

IN

OUT

WORDCLOCK