Understanding calibration
6.
Configure the densitometer to factory defaults and save the settings. (See the
densitometer manual for details.)
Factory default settings for the X-Rite DTP32 are:
7.
Turn on the PowerBook.
8.
In the PowerBook Setup Control Panel, select Normal for Modem Controls.
9.
Calibrate the densitometer, as described below.
T
O
CALIBRATE
THE
X-R
ITE
DTP32:
1.
On the X-Rite DTP32, simultaneously press the two buttons marked MENU.
The words
MAIN MENU
appear in the display.
2.
Press the p1 key until p2 appears.
3.
Press the cal key.
CALIBRATION
appears in the display.
4.
Press the den key.
CALIBRATING TRANSMISSION
is displayed, followed by the words
INSERT CAL STRIP
.
5.
Position the strip guide at 5.
6.
Insert the arrow end of the Auto-Cal strip into the entrance of the X-Rite DTP32 until
the roller starts pulling the strip.
Page 1
LANG (ENG)
TONE (loud)
TYPE (ser)
Page 2
BAUD (9600)
HAND (xon)
AXMT (on)
Page 3
DPT (on)
SEP (spc)
DLIM (crlf )
Page 4
DEF (off )
X10 (off )
DAP (off )
Page 5
M/M (max)
LOCK (off )