LF2020/LF2090 Sit-To-Stand Patient Lift Owner’s Manual

12

CHECKING THE BATTERIES

1. Check batteries by pressing the Battery

Testing Button (blue circle with a battery

sign) on the control box.

2. Batteries are fully charged when all lights on

the Battery Indicator are "ON".

3. Charge batteries when Indicator shown only

one “GREEN” light.

4. Do not use the lift when no “GREEN” light is

shown. Charge the batteries immediately.

5. Replace batteries when frequent charging

is needed. (Reorder #: DPL650BAT)

CHARGING THE BATTERIES

1. Ensure the power is switched "ON" (the red

“RESET” button should be up).

2. Insert charging plug into charging port on the

control box.

3. Plug charger to power supply.

4. All lights of battery indicator should be

"ON" while charging.

5. It takes approximately 2-3 hours to fully

charge the batteries from one green light.

6. It takes approximately 7-9 hours to fully

charge the batteries from the red light.

7. Do not leave the batteries on charging for an

extended period of time. This will shorten

the life of the batteries.

8. Do not let the batteries run down to the last

red light. This could shorten the life of the

batteries or damage the batteries.

9. Unplug the charger first before using the lift.

THE LF2020/LF2090 SIT-TO-STAND PATIENT

LIFT WILL NOT OPERATE WHILE CHARGING.

IMPORTANT NOTES ON CHARGING

1. Push lift to charging location and charge the

batteries with the charger provided. Avoid

unplugging hand control and motor from

control box. Frequent plugging/unplugging of

the hand control and motor into/from the

control box could damage the control box.

2. Do not charge batteries over 12 hours.

Red

Amber

Green

Green

Green

ON

ON

ON

ON

ON

FULLY CHARGED

ON

ON

ON

OFF

OFF

NEED CHARGE

ON

ON/OFF

OFF

OFF

OFF

CHARGE IMMEDIATELY. DON’T USE THE LIFT

OFF

OFF

OFF

OFF

OFF

BAD BATTERIES OR BAD CONNECTION

Know What The Battery Indicator Lights Are Telling You

Warning Buzzers
The control box has a buzzer that provides you with information on the status of the batteries and

the lift. To stop the buzzer, push down the RESET button or press the Battery Indicator.

BUZZER TYPE

INDICATION

NOTE

2 beeps/sec

Low Battery (red / amber lights on)

Charge batteries immediately

3 beeps/sec

Low Battery (red light on)

Charge batteries immediately

6 beeps/sec

Very Low Battery (no light on)

Charge batteries immediately

3 beeps every

other second

Bad Battery (no light on)

Replace batteries

3 long beeps

Overloaded. Batteries are charged

but lift does not lift.

Exceeded maximum load capacity