The temperature and relative humidity settings displayed for ClimateLine or ClimateLine

MAX

are not

measured values.

•

Check with your clinician/service provider before using the SlimLine air tubing with devices other
than the S9 or H5i.

Humidifier performance

The following settings have been tested at 22°C ambient temperature:

Cleanable water tub

Injection molded plastic, stainless steel and silicone seal

Standard water tub

Injection molded plastic, aluminium and thermoplastic
elastomer

Operating temperature

+5ºC to +35ºC

Operating humidity

10–95% non-condensing

Storage and transport temperature

-20ºC to +60ºC

Storage and transport humidity

10–95% non-condensing

ClimateLine air tubing

Flexible plastic and electrical components, 2 m, 15 mm
inner diameter

ClimateLine

MAX

air tubing

Flexible plastic and electrical components,1.9 m, 19 mm
inner diameter

Heated tubing temperature cut-out

 41°C

IEC 60601-1 classification

Class II (double insulation), Type BF

Electromagnetic compatibility

Please refer to the S9 Information Guide for details

CPAP mask

pressure, cm H

2

O

RH output %

Nominal system output AH

a

, BTPS

b

a.

AH - Absolute Humidity in mg/L.

b. BTPS - Body Temperature Pressure Saturated.

Setting 3

Setting 6

Setting 3

Setting 6

4

90

100

10

18

10

95

100

11.5

21

20

95

100

11

18

25

100

100

12

13.5