SECTION 6—OPERATING INSTRUCTIONS
Part No 1148112
29
Connecting/Positioning the Nasal
Cannula
CAUTION
To ensure proper breath detection and oxygen delivery, DO
NOT use tubing length exceeding 25 ft (7.6 m).
NOTE: For this procedure, refer to FIGURE 6.7.
NOTE: Invacare recommends using the XPO
2
with a 4 ft (1.2 m)
cannula. The tubing must be connected to the XPO
2
’s oxygen outlet
as shown below.
NOTE: Replace the nasal cannula on a regular basis. Check with
your equipment provider or physician to determine how often the
cannula should be replaced.
NOTE: DO NOT share cannulas between patients.
1. Connect the nasal cannula to the XPO
2
’s oxygen outlet port
(Detail “A”).
2. Place the cannula over your ears and position the prongs
in your nose as instructed by your health care provider or
cannula manufacturer (Detail “B”).
FIGURE 6.7 Connecting/Positioning the Nasal Cannula
Nasal
Cannula
Oxygen
Outlet Port
DETAIL
“A”
DETAIL “B”