OPERATING MANUAL

REMOTE OPERATION

Release 2.0 (98/06)

4-3

4.2.3

Rear Panel J2 Connector (continued)

See Table 4.2-2 for the

list of the J2 connector terminal numbers, their labels, and corresponding functions.

Table 4.2-2 Rear Panel J2 Connector Terminals and Functions

Connector

Label

Name

Function

J2-1

VRMTSELECT

Remote Output Voltage

Programming Select

Selects remote output voltage
programming when jumpered to pin 3.

J2-2

IRMTSELECT

Remote Output Current Limit

Programming Select

Selects remote output current limit
programming when jumpered to pin 3.

J2-3

CONTROL
GND

Control Ground

Control ground.

J2-4

N/C

No connection

None.

J2-5

VPGMRTN

Voltage Program Signal Return

Return for voltage program signal.

J2-6

VPGM

Output Voltage Programming
Input

Input for voltage programming signals from
an analog device.

J2-7

IPGMRTN

Current Program Signal Return

Return for current program signal.

J2-8

IPGM

Output Current Limit
Programming Input

Input for current limit programming signals
from an analog device.

J2-9

VMONRTN

Voltage Monitor Signal Return

Return for voltage monitor signal.

J2-10

VMON

Output Voltage Monitor

Output for output voltage monitor signal.

J2-11

IMONRTN

Current Monitor Signal Return

Return for current monitor signal.

J2-12

IMON

Output Current Monitor

Output for output current monitor signal.

J2-13

N/C

No connection

None.

J2-14

TTL S/D RTN

TTL Shutdown Signal Return (-)

Return for TTL shutdown signal.

J2-15

TTL S/D

TTL S/D Input (+)

Input for TTL shutdown signal.

Making J2 Connections

CAUTION

Do not attempt to bias program/monitor signal return (J2 terminals 5, 7, 9,
and 11) relative to the supply output because control ground (J2-3) and the
program/monitor signal returns are at the same potential as the power
supply return in a standard unit. Use the Isolated Programming (ISOL)
Interface option to allow control from a programming source at a different
potential relative to the supply’s output.

Make connections to the J2 connector using its screw-type wire clamps. Before making any connections, turn the
power supply OFF and wait until the front panel displays have gone out. You can unplug the connector from the
back of the unit in order to make it easier to install the required wiring.

Wiring

Use any suitable wire such as 16 to 24 AWG stranded wire for most connections, but we recommend that you
use shielded twisted pair wiring for applications which require low noise performance. Strip wires 0.26"
(6.5 mm) and insert securely as with any wire clamp connector.