Page 40: Operation

Advertising
background image

22

Rev. 5

Operation

Energizing Power Supply System Components

If this occurs, turn the Model 430 Programmer off, wait 15 seconds

or more, and power the Model 430 Programmer back on.

When powering the system off, first turn off the power supply controlled by

the Model 430 Programmer followed by the Model 430 Programmer. The

controller will then ensure the load sees no abnormal power transients as

the power supply is turning off.

3.1.2 Energizing

Power

Supply

Warning

Do not change power supply jumpers, dip-switches, or other factory

settings. If not rack-mounted, always position power supplies for

convenience in disconnecting the power cords.

Caution

The master unit must be turned on between 2 to 5 seconds before the

slave, otherwise the units will fault and the power-up sequence will

have to be repeated.

Place the power supply switches in the ON position in sequence

(see Caution Note). No local (front panel) adjustments or

connections are required since the power supply control mode and

other parameters have been factory-configured for control by the

AMI Model 430 Power Supply Programmer.

When powering the system off, turn OFF the power supplies

(master unit last) before powering off the Model 430 Programmer.

3.2

Model 430 Programmer Default Display

The default display is illustrated in the figure below. It is displayed

whenever no menus are being accessed and no errors are being indicated.

The default display can be thought of as being logically divided into four