6 special configurations, 8 short-circuit or resistive load, Installation – American Magnetics 4Q06125PS-430 High Stability Integrated Power Supply System User Manual

Page 34

Advertising
background image

16

Rev.

3

Installation

Magnets w/o Persistent Switch

Aux connector

J2

on the rear panel of the Model 13x Liquid Helium

Level Instrument (12). Refer to section A.6.2 on page 151.

i. Connect each device line cord from the respective device to the

appropriate power receptacle.

j. Remote communications via Ethernet and/or RS-232 can be

accomplished by connecting suitable cabling to the Model 430

Programmer rear panel

ETHERNET

and/or

RS-232

connectors.

2.6 Special Configurations

The Model 430 Programmer has been designed for optimal operation with

a superconducting magnet (i.e. a very low resistance, high inductance

load) with a persistent switch. The Model 430 Programmer is capable of

controlling current to other loads; however, some modification to the

Model 430 Programmer settings and/or connections must usually be made.

Two commonly encountered configurations are: 1) superconducting

magnets without a persistent switch, and 2) operation on a short-circuit or

low resistance load.

2.7 Superconducting Magnets with No Persistent Switch

An external stabilizing resistor for superconducting magnets without a

persistent switch is no longer required

1

. However, these systems do

require a specific Model 430 Programmer stability setting based on the

magnet inductance as follows:

For magnet inductance <= 100 Henries (H):

Stability Setting = (100 - H)

For magnet inductance > 100 Henries:

Stability Setting = 0

2.8 Short-Circuit or Resistive Load

If operating with a short-circuit as a load without the presence of a

superconducting magnet, the Model 430 Programmer must be manually

configured for stability. Normally, when the persistent switch heater is

deactivated, the Model 430 Programmer sees essentially a short-circuit

load since the persistent switch shunts all current flow away from any

connected magnet. Therefore, one method of operating a short-circuit is to

1. Effective with Model 430 firmware version 1.62.

Advertising