Technical data – M&C TechGroup PSP4000-H_C_T Operator's manual User Manual

Page 8

Advertising
background image

8

Gas sampling and gas conditioning technology

2-2.1-ME

9

TECHNICAL DATA

Probe

series SP®

portable version
PSP4000

PSP4000-H

PSP4000-H / C

PSP4000-H /C /T

Part No. 230V

40S1000

40S1005

40S1015

Part No. 115V

40S1000a

40S1005a

40S1015a

Sample temperature

max. 600 °C *standard

Sample pressure

max. 1 bar

Ambient temperature

-20 °C to +60 °C

Filtration chamber
volume

40 cm

3

Filter element

S-2K ceramic

, 2 μm

Probe temperature

Adjustable between 100-180 °C, pre-set at works to 180°C

Ready for operation

After approximately 30 minutes

Gas inlet

Basic connection G 3/8"i, probe tube or adapters optional

Gas outlet

1/8" NPT + tube connector ø 6 mm (optional ø 8 mm)
and pipe clamp for attachment of heated sample line

Power supply

220-240V 50/60Hz, 200 W or 115V 60Hz

Electrical connection

Plug and socket connector 7 pin with 4 meter connection cable

Electrical equipment
standard

EN 61010, EN 60519-1

System of protection

IP40, EN 60529

Material

Stainless steel 316Ti, ceramic, FPM

Weight

3 kg

Calibration gas
connection for
SS-tube/Plastic
hose ø 6 mm

no

yes

yes

Temperature
measurement with
Thermocouple sensor
FeCuNi, length 600 mm
with 4 m connection
cable and standard plug.

no

no

yes