OPERATING OF THE BUILT-IN ELECTRONIC TEMPERATURE CONTROLLER

N O T E !

Only use temperature controller 01 B 8365.

In normal operation the display of the temperature controller shows the actual cooling temperature.
Figure 7 shows the front view of the temperature controller.

Figure 7 Front view of the temperature controller


19.1

CHANGING THE SET VALUE

To change the set value the P-key has to be pushed < 2 sec. The company fixed value of 5°C
appears. With the two arrow keys the value can be changed. This value should not be adjusted lower
than +1°C, because otherwise the heat exchanger will freeze up. Is the value adjusted higher than the
ambient temperature the cooler will not work.

19.2

CHANGING THE TEMPERATURE ALARM WINDOW

The exact description of changing the temperature alarm window can be found in the manual of the
temperature controller in the appendix.