RECEPTION AND STORAGE

The EC/ECS gas cooler is a complete pre-installed unit.

Carefully inspect the EC/ECS and any special accessories included with it immediately on arrival

by removing them from the packing and checking for missing articles against the packing list!

Check the items for any damage in transit and, if required, inform the shipping insurance company

immediately of the damage found!

N O T E !

The cooler must be stored in a weather-protected frost-free area!

N O T E !

During transport and when in storage, the cooler must always be stood up
with the transport feet positioned underneath so that the oil in the closed
compressor circuit cannot run out of this compressor case. If the cooler is
transported on its back by mistake, it must stood in the operating position
for approx. 24 hours before it is switched on!



12

INSTALLATION INSTRUCTIONS

The EC/ECS cooler is equally suitable for wall mounting or for installation in a 19" rack.

N O T E !


Please state the desired type of mounting when ordering so that the LED
function display can be positioned to match at the factory!

N O T E !

The operating position for this cooler is exclusively vertical. This is the only
way to ensure proper separation and removal of condensate in the heat
exchangers. During transport and installation, the cooler must always be
stood up with the transport feet positioned underneath so that the oil in the
closed compressor circuit cannot run out of the compressor case.
The cooler should be kept away from sources of heat and well ventilated
when installed, so that condensation from warmth will not occur and
interfere with operation.
The minimum installation dimensions (fig. 2) must be followed without fail. If
the unit is installed in the open, the cooler must be installed in a housing
that is frost-free in winter and adequately ventilated in summer. Avoid
locating the unit in direct sunlight.

Unheated gas sample lines must be provided with slope up to the cooler. In
that case pre-separation of the condensate is not required.

Connect the heated sample line with sufficient thermal decoupling to the
cooler!