24

Gas sampling and gas conditioning technology

3-4.1-ME

LED display Function error

and status

alarm

Probable cause

Checking / Correction

°C >

ON

°C <

green

red

Equipment does
not cool;



Cooler has been
over-cooled (temp.
< 2°C);

Ambient
temperature < 2°C



Cooling
compressor
stopped;




Cooling compres-
sor runs continu-
ously;

Ambient temperature must be ≥ 8°C!



Check temperature at EC automatic control board
(18.2);
If temperature < 2°C (< 0,2V):
Check PT100 temperature sensor (19);
If not OK:
Replace sensor.


Solid-state relay defective;
Replace EC automatic control board;