Electrical connections, Figure 4, Electrical sockets – M&C TechGroup EC-EX Operator's manual User Manual

Page 15: 2 electrical connections

Advertising
background image

15

Gas sampling and gas conditioning technology

3-4.1ME EC-EX 9904

15.2

ELECTRICAL CONNECTIONS

W A R N I N G !

When connecting the equipment, please ensure that the supply
voltage is identical with the information provided on the model type
plate.

N O T E !

Attention must be paid to the requirements of IEC 364 (DIN VDE
0100) when setting high-power electrical units with nominal
voltages of up to 1000 V, together with the associated standards
and stipulations.

Check the details on the type plate to ensure that the equipment is
connected up to the correct mains voltage.

The main circuit must be equipped with a fuse of 10A

T

for the 230V

version and 16A

T

for the 115V version (automatic cut-out); for

electrical details see technical data.


Figure 4 shows the electrical connections at the plastic housing behind the front panel of the EC-EX
casing (Fig. 2).

Figure 4 Electrical sockets

Power supply : 230V/50Hz oder 115V/50-60Hz (see type plate)
Status alarm : one changeover contact
Contact rating : 230V AC, 2A, 100VA or
230V DC, 2A, 50W

Power
supply

1 L
2 N

3 PE

4

230V 50Hz

5
6
7
8

11 NC

12 Com

13 NO

Alarm
contact

230V 100VA

Advertising