Temperature setting for the ec/ecs cooler, Figure 6, Temperature setting – M&C TechGroup EC-EX Operator's manual User Manual

Page 23

Advertising
background image

23

Gas sampling and gas conditioning technology

3-4.1ME EC-EX 9904

20.1

TEMPERATURE SETTING FOR THE EC/ECS COOLER

The EC-EX gas cooler is set at the factory to a regulated temperature of +5°C.

Setting of the regulated temperature is done by trimming potentiometer P1, on the EC-EX automatic
control board of the cooler. The setting range covers from 0°C to 20°C. Turning it to the left sets a
lower temperature, and turning it to the right sets a higher temperature.


















Figure 6 Temperature setting



Connecting an external temperature measuring device into the cooling block out of aluminium the real
temperature can be taken and controlled.

N O T E !

Do not adjust the temperature below +2°C or above +7°C.
Freezing-up of the heat exchanger can happen below +2°C.
Above +7°C the cooling unit will not work sufficiently!



-

+