Technical data – M&C TechGroup FT-...-H2 Operator's manual User Manual
Page 8

8
Gas sampling and gas conditioning technology
5-2.1.1-ME
7
TECHNICAL DATA
Filter series FT
®
FT-3G-H2
FT-3SS-H2
FT-0,1GF-H2
Part number
01 F 3006(a)
01 F 3008(a)
01 F 3007(a)
Filter finess/material of filter element
3µm / glass
3µm / stainl. steel
0,1µm / glassfibre
Power supply
230 - 240V 50Hz, 350VA or 115V 60Hz (a)
Electrical connection
2,5mm
2
terminals, 2x PG13 cable glands
Protection classification/electrical
standard
IP44-EN60529 / EN61010, EN60519-1
Ambient temperature max.
+50°C
Sample temperature max.
+180°C
Ready for operation
after approx. 30 minutes
Temperature alarm contact
alarm point
T 30°C at T
set
, contact rating 250V, 3A, 0,25A=
Regulation of temperature
thermostat (PT100 as option)
Sample connections
G1/4” i
(DIN ISO 228/1)
Stagnant space
65cm³
Filter surface
70cm³
Ma. pressure
4 bar g
Dimensions (w x h x d) / weight
350 x 300 x 140mm / 6kg
Method of mounting
wall mounting
Material of sample contacting parts
PTFE, glass, FPM
(a) 115V-Version