Cleaningtype n 3/5/9 kpe – M&C TechGroup CSS-VC2 Operator's manual User Manual
Page 45

14.1a ME
Gas sampling and gas conditioning technology
45
Check ease of movement of the pump by turning the fan wheel.
Move structure diaphragm via fan wheel into the top dead centre (dead point) position.
Now hand tighten the screws 4.
18.2.5 CLEANINGTYPE N 3/5/9 KPE
On changing valve plates and diaphragm, prior to assembly of the pump head, all parts should be
checked for contamination and cleaned if necessary.
Wipe the parts as dry as possible with a cloth. Solvents should not be used when cleaning
because they may damage the plastic parts. If oil-free compressed air is available, blow out the
parts.
W A R N I N G !
Aggressive media residues possible.
On dismantling, repairing or cleaning the pump, protective
glasses and protective clothing must be worn!
18.3
MAINTENANCE OF THE BUILT-IN PERISTALTIC PUMP TYPE SR 25.2
Pump hose, belt, pressure rollers and springs are the only wear parts of the pump. They are easy to
change.