M&C TechGroup CSS-VC2 Operator's manual User Manual
Page 53

14.1a ME
Gas sampling and gas conditioning technology
53
Messages that may appear in the list of events and their meanings:
Alarms
Event message
Significance
Everything OK
Everything OK
T1 alarm
Cooler temp. too high
T2 alarm
Temp. channel 2 is too high
T3 alarm
Temp. channel 3 is too high
T1 low alarm
T1 under the alarm hysteresis
LA1 cable br.
Cable break humidity sensor 1
LA2 cable br.
Cable break humidity sensor 2
LA1 alarm
Alarm humidity sensor 1
LA2 alarm
Alarm humidity sensor 2
T1-temp> max
T1> 99.9° C => possible cable break
T2-temp> max
T1> 248 ° C => possible cable break
T3 temp> max
T1> 248 ° C => possible cable break
Th2 Error
If thermocouple is connected and temperature does not rise => defective
thermocouple
Th3 Error
If thermocouple is connected and temperature does not rise => thermocouple
is faulty
T1-Temp <min
T1 <-70 ° C => possible short circuit
T2-Temp <min
T2 <-50 ° C => possible short circuit
T3-Temp <min
T3 <-50 ° C => possible short circuit
Warnings
Event message
Significance
T2 low warning
T2 low temp. (outside of hysteresis)
T3 low warning
T3 low temp. (outside of hysteresis)
Unit cold
Device temp. too low: <1° C
Device hot
Device temp. too high: > 50° C
FA1 warning
Flow alarm channel 1
FA2 warning
Flow warning channel 2
Pump1 warning.
Pump1 out, caused by an alarm (fluid/temperature)
Pump2 warning.
Pump2 out, caused by an alarm (fluid/temperature)
Calibration
defective
Internal error triggers service interval and requires return of the unit
Service int.
Service interval expired
Flushing pressure
off
Flushing pressure defective