External Sealing
There are two different methods for external sealing, a crimped wire seal, and the application of a tamper proof label.
Wire Sealing
Remove the two flat head screws from one side of the
Indicator housing. Replace these screws using the two
security screws from the Weights and Measures kit. Using
the crimp seal and wire, seal the Indicator as shown in
Figure B-4.
Tamper Proof Label Sealing
Place an adhesive seal over one of the flat head screws at the
side of the Indicator housing. An alternative larger label seal,
can be used to cover one or more of the side screws and the
housing edges. See Figure B-5.
Figure B-4. External Wire Seal.
Figure B-5. External Sealing Label.
ALTERNATE LABEL
DETAIL
FRONT HOUSING
SEALING LABEL
SCREW
REAR HOUSING
SEALING LABEL