14 printing data – Ohaus CD-11 Indicator Manual multi User Manual
Page 38

EN-32
CD-11 Indicator
4.14 Printing Data
Printing data to an external computer or printer requires that the communication parameters in the Print menu be
set first. Refer to paragraph 2.6.6 Print Menu for proper set up.
To print data, press the PRINT/
Units button. The display acknowledges by momentarily blinking off.
NOTE: If the PRINT/
Units button is held down too long, the display will advance to another measuring unit
without sending the print data.
Command
Character
Description
?
Print current mode: kg, g, lb, oz.
P
Same as pressing PRINT button.
T
Same as pressing TARE button.
Z
Same as pressing ZERO button.
xS
Print Stable only. Where x=0 Off, and x=1 On
A S
A S
A S
A S
A S
Automatically send data when stable after motion.
xxxxS
Send at interval. Where xxxx=1 to 3600 seconds.
CS
Send as fast as possible (continuous print)
M
Increment to next enabled unit
RS232 USER COMMAND TABLE
4.13 RS232 Commands (Cont.)
NOTE: To turn auto printing, interval printing or continuous printing off, send 1S or 0S to reset normal printing
mode.