EN-32

CD-11 Indicator

4.14 Printing Data

Printing data to an external computer or printer requires that the communication parameters in the Print menu be
set first. Refer to paragraph 2.6.6 Print Menu for proper set up.

To print data, press the PRINT/

Units button. The display acknowledges by momentarily blinking off.

NOTE: If the PRINT/

Units button is held down too long, the display will advance to another measuring unit

without sending the print data.

Command

Character

Description

?

Print current mode: kg, g, lb, oz.

P

Same as pressing PRINT button.

T

Same as pressing TARE button.

Z

Same as pressing ZERO button.

xS

Print Stable only. Where x=0 Off, and x=1 On

A S

A S

A S

A S

A S

Automatically send data when stable after motion.

xxxxS

Send at interval. Where xxxx=1 to 3600 seconds.

CS

Send as fast as possible (continuous print)

M

Increment to next enabled unit

RS232 USER COMMAND TABLE

4.13 RS232 Commands (Cont.)

NOTE: To turn auto printing, interval printing or continuous printing off, send 1S or 0S to reset normal printing
mode.