P/N 80103490 C © Ohaus Corporation 2013, all rights reserved.

Printed in USA

Certificate of Conformance 05-118 has been issued by the National Conference on Weights and Measures, confirming
that the Adventurer Pro models listed in this manual conform to the requirements in NCWM Publication 14 and NIST
Handbook 44.

Before this balance can be used in commercial trade, it must be inspected by local weights and measure authorities.
It is the responsibility of the purchaser to ensure that all pertinent legal requirements are met. Please contact your local
weights and measures office for further details.

Item Number

AV64N AV64CN

AV114N AV114CN

AV264N AV264CN

AV213N AV213CN

AV313N AV313CN

AV413N AV413CN

AV2102N AV2102CN

AV3102N AV3102CN

AV4102N AV4102CN

Max (g)

65

110

260

210

310

410

2100

3100

4100

Min (g)

0.1

0.1

0.1

0.2

0.2

0.2

5

5

5

d=(g)

0.0001

0.0001

0.0001

0.001

0.001

0.001

0.01

0.01

0.01

e=(g)

0.001

0.001

0.001

0.01

0.01

0.01

0.1

0.1

0.1

Class

I

I

I

II

II

II

II

II

II

Repeatability (g)

0.0001

0.0001

0.0001

0.001

0.001

0.001

0.01

0.01

0.01

Linearity (g)

+

0.0002

+

0.0002

+

0.0003

+

0.002

+

0.002

+

0.002

+

0.02

+

0.02

+

0.02

Weighing Units

mg, g, ct, dwt, oz, ozt, GN

g, ct, dwt, oz, ozt, lb

g, kg, dwt, oz, ozt, lb

Application Modes

Weighing, Animal Weighing, Parts Counting, Percent Weighing, Check Weighing, Totalize

Tare Range

To capacity by subtraction

Stabilization Time

3

Power Requirements

AC Adapter (included)

Calibration

Digital with external weight

Display Type

2-line LCD with Backlight

Display Size (mm / in)

100 x 25 / 4 x 1

Pan Size (mm / in)

90 / 3.5

120 / 4.7

168 x 180 / 6.6 x 7.1

Dimensions WxDxH (mm / in)

220 x 300 x 300 8.7 x 12 x 11.8

220 x 190 x 300 8.7 x 7.5 x 11.8

220 x 85 x 300 8.7 x 3.3 x 11.8

Net Weight (kg / lb)

4.0 / 8.7

3.3 / 7.3

2.8 / 6.1

Net Weight (with InCal) (kg / lb)

4.4 / 9.6

3.7 / 8.2

3.2 / 6.9

AV...CN models are equipped with internal calibration.

Introduction

This document is a supplement to the Instruction Manual provided with the Adventurer Pro balance. It is intended to
provide information specific to the NTEP Certified models. For all other information, refer to the Instruction Manual.

Setting the Balance

Press and hold the

Menu-Cal button until MENU is displayed. Move to the Setup Sub-menu using the No button.

When SETUP is displayed, press the

Yes button to enter the Setup Sub-menu. When LEGAL TRADE is displayed,

press the

Yes button to view the setting OFF. Press the Yes button to change the setting to NTEP. Press the Yes but-

ton to store the setting. Press the

Exit button to return to the current weighing mode.

Note: Before setting the balance, enter the Menu to perform a calibration and make any desired balance setting

adjustments. Once the balance is set, the following conditions apply:

• The Calibrate Sub-menu is hidden and not available.

• In the Setup Sub-menu, the Legal Trade menu item is locked.

• In the Setup Sub-menu, the Auto Zero (AZSM) menu item is set to 0.5d and locked.

• In the Readout Sub-menu, the Gross Indicator menu item setting is locked.

• The Mode Sub-menu is locked.

• In the Unit Sub-menu, the Unit menu items are restricted to Systems International units.

• In the Print Sub-menu, the Output When Stable menu item is set to ON and locked.

• In the Print Sub-menu, the Continuous setting in the Auto Print menu item is hidden and not available.

• In the Reset Sub-menu, the Reset Readout, Reset Mode, Reset Unit and Reset Global menu items are

locked.

• In the Lockout Sub-menu, the Calibration menu item is locked.

To regain access to the locked Menu settings, turn the balance off. Turn the balance on and press the Lock Switch
while OHAUS is displayed during power up.

Sealing the Balance

To seal the balance, cover the Lock Switch access hole by sliding the Security Cover to the right. Then install a
wire through the holes in the tabs of the Security cover and Top Housing.

Security Cover

Wire Seal