Mmode, Percnt – Ohaus DEFENDER 7000 BENCH SCALES Manual en User Manual

Page 52

Advertising
background image

EN-50 7000 Series Indicators

1 SECOND

2 SECONDS

3 SECONDS

4 SECONDS

mMOdE

PERCENT

mMOdE

DYNAMIC

*

0

0.000

kg

ENTER REF

PErCNt

%

CLEAR REF?

*

0

0.000

kg

PLACE REF OR

100.0

%

12.345 KG

*

0

12.345

kg

PLACE REF OR

12.345

kg

ENTER REF

*

0

0.000

kg

READY

*

12.345

kg

5 SECONDS

*

12.345

kg

HOLD

4.9.3

Percent Weighing

Use this mode to compare the weight of items as a percentage of a Reference Weight.

Defining the Reference Weight
When the Mode button is released, CLEAR REF? is displayed. Press No to use the stored Reference
Weight or press Yes to establish a new Reference Weight.

When establishing a new Reference Weight., the display alternates between PLACE REF OR
and ENTER REF.



To establish the Reference Weight using an item, place the item on the scale and press FUNCTION.



To establish the Reference Weight using a numeric value, enter the value using the keypad,
then press FUNCTION.


Percent Weighing
After defining the reference weight, place an item on the scale and read the value on the display.
The percent value is displayed on the top line and the actual weight is displayed on the bottom line.

NOTE: Press FUNCTION to briefly display the Reference Weight.

4.9.4

Dynamic Weighing

Use this mode to weigh moving items or large items that block the display.

NOTE: If Accumulate is set to AUTO in the Setup menu, the held weight is automatically added to the
accumulation data.

Manual operation (DYNAMIC is set to MANUAL in the MODE menu)
When the display shows READY, place the item on the scale.


Press FUNCTION to average the readings for the time period defined in the LEVEL setting
(Section 3.6.6).








When averaging is completed, the dynamic icon flashes. The averaged weight and HOLD are
displayed until FUNCTION is pressed again.



Semi-automatic operation (DYNAMIC is set to SEMI in the MODE menu)
When the display shows READY, place the item on the scale. The readings are automatically averaged
for the time period defined in the LEVEL setting. The averaged weight and HOLD are displayed until the
item is removed and FUNCTION is pressed.

Automatic operation (DYNAMIC is set to AUTOMATIC in the MODE menu)
When the display shows READY, place the item on the scale. The readings are automatically averaged
for the time period defined in the LEVEL setting. The averaged weight and HOLD are displayed until the
item has been removed from the scale for 10 seconds.

Advertising