Specific Gravity Determination

Specific Gravity Determination

Specific Gravity Determination

Specific Gravity Determination

Specific Gravity Determination

A Clamp and Rod Support, Ohaus part number 183-00, is an accessory used to
elevate the Balance for suspending a specimen in water. The rod is inserted into
the 1/2 inch recess on the underside of the base. By means of a fine wire or thread,
the specimen is attached to the check pin which connects the platform loop to the
check rod. The procedure for weighing a suspended specimen is the same as that
for weighing an unknown mass on the Platform.

CARE AND MAINTENANCE

CARE AND MAINTENANCE

CARE AND MAINTENANCE

CARE AND MAINTENANCE

CARE AND MAINTENANCE

The following practices will assure that your Balance will give you years of
satisfactory and trouble-free service.

When the Balance is not in use, remove any mass or weights from the Platforms.
Move the Tare Poise from zero so that the Balance is not in equilibrium. This will
stop the Balance from oscillating and prevent unnecessary wear.

Use and store the Balance in places free of vibration. Never

Never

Never

Never

Never lubricate the scale

bearings. The bearings in these Balances are high grade polished agate V-blocks
and the knife edges are hardened, precision-ground steel. This type of bearing
works most efficiently when clean and dry. Keep the Balance clean at all times,
being particularly careful not to let dirt accumulate in the vicinity of the bearings. If
the bearings should become dirty, try to clean them using an air syringe. The Model
106-00 Balance Cover is recommended to keep the Balance clean when not in use.

Cleaning Magnet Faces

Cleaning Magnet Faces

Cleaning Magnet Faces

Cleaning Magnet Faces

Cleaning Magnet Faces

In general, most foreign matter may be easily removed by an air syringe. If the
damper vane still does not move freely, proceed as follows:

1. CAREFULLY

CAREFULLY

CAREFULLY

CAREFULLY

CAREFULLY turn the Balance on its side.

2. Remove the four (4) screws that mount the magnet case to the underside of

the base.

3. Remove the magnet case.

4. Clean the magnet faces by pressing a piece of adhesive-backed tape against

them.

5. Reinstall the magnet case.